Calibration of a Tektronix 5113 Oscilloscope

Ebay, here we go again

I had only the chance to buy it on Ebay. Well the price was not high, and three modules were with it. It was said to work very good, and not dead on arrival. Well it worked for a few hours, and then the CRT stopped. The High Voltage transformer was whistling badly, and I thought it was bad. I was lucky, it was only the first capacitor was shorted. At 10nF, 4kV these are a bit rare, but I found a replacement. The original was a disc type, which is sintered material. Frankly, at 3400V DC, such should not be used. Such "isolated" crystals inside, it's just asking for trouble. On the Tektronix forum, they already wrote, it's probably NOT the transformer, and they were right. So the replacement is a foil cap now.

After the HV was working again, it appeared the Channel chopper oscillator was dead. The lousy seller must have known it, because he removed the chopping blanking transistor. (Leave it in, and the screen blanks when you choose "chop" but there is no chopper frequency). So he nicely frauded this sale. This man was a school teacher in Holland. He was too stupid to find the cause of the failure, and so he sets up this fraud, to sells the scope "as is". Oh well, same as politicians are always lying.

But on the other hand, it is satisfactory to find out WHY the item was sold, if you can fix it nicely. Then chances on any other nasty problems are lower.

Disclaimer.

You will screw things up when changing the setting of your time base module experimentally, like learning by doing. It's not going to be like that, if one calibration depends on another first! In that case, you need the right order. So you can change the vertical amplifier sensitivity, but that provides you will not change the CRT high voltage after that. If you change the High Voltage (3400V) of the CRT tube, you need to re calibrate everything related to this, like horizontal and vertical amplifier sensitivity, focus, intensity, and whatever else is related. So just try to see what happens is not a good idea with the time base setting. Probably the easy to do settings were maintained already by previous owners. But the harder they are to do, the less chances are they were done at all. So now, after 50 years not calibrating the horizontal amplifier, and the time base module, it's time to do those harder settings now.

However this is no guaranteed guidance for how to do a calibration yourself. On the other hand, such calibrations of which most people know they should not touch it, like horizontal and vertical amplifiers, after 50 years are not really accurate any more. So going through some procedures with those, may pay off. The following is only I did this myself, bypassing the need of special calibration tools, and it is no calibration guide line in general.

Where I get my information from:

  1. Tekwiki for the documentation. (you can participate there, very friendly website owner)
  2. The Tek Scopes forum for when you want to discuss. This is one of the very few forums I have seen, where you get a normal answer from normal people. No dominant inmates there, disciplining the newcomers. Many retired Tek engineers there, soldering on their beloved items, being really helpful to others.

The use of a time mark generator.

The Tektronix manuals always write, take a time mark generator, take a voltage calibrator, and take a calibrated vertical module. Take this and that that.... But hey... I don't have all these items to "take".

So I thought, today's very accurate multi meters and universal signal generators should make it possible perhaps. I asked on the Tektronix forum if this is really needed, but they said that touching the time base calibration, without time mark generator, is something I would probably regret. I did find a work around however.

At first, I followed the Tektronix official calibration routine by the book, just to see what could be done without special tools, and to see where it gets stuck. Now there is some point in this routine, where I would say it's the chicken or the egg. They say "take a calibrated vertical module". And then, with that they adjust the vertical amplifier. Mind you, this is the official TEK procedure. This is no good way! The result of that calibration depends on how well this vertical amplifier module which you just "take", was calibrated. In my case: Not calibrated at all. My conclusion: Forget this. And I suppose this was what the guys in the forum were trying to tell me.

Here is how I did the calibration still

Equipment used:

5113 Dual Beam, Dual Storage Oscilloscope
5A14N vertical module in slot #1.
5B10 Time base in slot #3
Agilent 34401A multi meter
Agilent 33120A Arbitrary Waveform Generator (= very versatile Oscillator)

If not calibrating the horizontal and vertical amplifiers first, you get a mess, when the time base is 5% too slow, and you would just set the horizontal amplifier 5% higher gain. Because that would also make the beam move faster.

So, calibrating these mainframe scopes, means the horizontal and vertical amplifiers have to be calibrated first, and after that, the time base module and vertical modules can be calibrated, using this scope, with now calibrated horizontal and vertical amplifiers. That is the whole principle.

The sensitivity of the horizontal and vertical amplifiers is normalized at 50mV per division. This has to be set first, and directly like this. So no matter what the set up is, at 200mV at the green connectors inside the mainframe, the beam should move 4 divisions, both horizontal or vertical.

The question is, how to get this 200mV on the mainframe connector, so at the input of the horizontal and vertical amplifier? There are more ways for this, and in the TEK calibration instructions, they use a very difficult method, which needs of course again a time mark generator to "take" and a calibrator for the AC signals to "take".

I think when using an Agilent 33120A Arbitrary Waveform Generator, this is all we need, and frankly it works a lot faster, easier to understand, and with better result. I verified the output signals of the Agilent 33120A, with the Agilent 34401A multi meter, and both read exactly the same values and frequencies, with 4...5 digits precision, also at millivolt level.

Another pretty nice reference is the internal calibrator of the scope itself. It is just there, and no long wires needed. This thing is locked to the mains frequency, and it gives a square wave of 2x the mains frequency. We have 100Hz here, which is more convenient than in 60Hz regions. So it gives a perfect 0.1 Second time definition. However it has no good 50% duty cycle, so it can only be used to display 0.1 second precisely, but not 0.05 second (for half a cycle). The amplitude is very accurate, provided the amplitude is calibrated. A problem with this is, the duty cycle is not symmetric. So you can not measure the RMS voltage to calibrate it.

To calibrate it, first measure exactly the duty cycle of it, use that as a correction factor, and then you can calibrate it with the Agilent 34401A multi meter, or any other which does RMS volt measurement accurate. So suppose, the signal is 46% high, and 54% low, you have to calibrate it for 184mV RMS. (400*0,46). That would give you indeed an precise calibration signal, because the 400mV is derived from the +30V which is a very stabile circuit. The Agilent 34401A multi meter can do this very good.

However there is no calibration potmeter for that, and the only way is to change the amplitude is change the +30V supply. The +30V has a very small tolerance, which I used to calibrate the scope calibrator output. So I set the +30V until I had exactly the required value, and the +30V was still within calibration. Then I set the -30V to the same level. So +30 and -30 are nicely balanced.

As a second test, after calibrating the vertical amplifier of the scope, this calibration signal must test exactly at 400mV peal value. If not, one of the two actions was done wrong. So you have a double check on the vertical amp calibration, and on your work.

First CRT geometry must be perfect. Forget touching the Horizontal and Vertical amplifier calibrations, if the CRT picture is unlinear or has distorted edges. Go by the book, and do repairs if needed. If you can't get the geometry right, do not touch the Horizontal and Vertical amplifier calibration. Luckily with my 5113 that worked out excellent. So I could just continue from there. Since the CRT tends to distort at 5 divisions a lot more than at 3 or 4 divisions, I preferred to take 4 divisions.

The calibration process

This process is not complete. It is only how I calibrated some parts.

Preparations.

#1. The Tektronix method.

  • The Tektronix calibration routine at some point says_ "Take time mark generator, and take voltage calibrator". However I did not find them on my shelves.
  • Then, it comes to a point where you need to swap the time base module and the vertical module. As you can see in this picture.
  • This rotates what you see on the CRT by 90°, but it also swaps the connections of the horizontal and vertical amplifiers.
  • In some curious way, they work around the problem not forcing the user to measure directly on the pins 7A and 7B, which is the input of the vertical amplifier.
  • This process, says to use a vertical module which they say must be trusted. In other words: calibrated. But I stopped looking on my shelves. I don't seem to have it. Just calibrating the module is not possible without connecting wires to Pin 7A and 7B, or use a calibrated mainframe.
  • So one way or another, you get stuck in a chicken or the egg situation, and people on the TEK forum warned me for this.
  • However, I do not see it as a big problem for myself to connect two wires to pins 7A and 7B. Why not do so?!! See the picture for that.
  • And yes it worked, but I was not fully satisfied, because this process calibrates the horizontal amplifier by replicating the result of the vertical amplifier. Which replicates also the vertical geometry error, if there is one. And there will be one, but I have no way to find out, and it reduces the overall precision.
  • The process worked, but there were some small errors with the scope here and there. Though a lot better as it was before, it looked still like a not so well done calibration to me.
  • But then, if I do connect two wires there anyway, I can go for a much better process, not calibrating the one deflection amplifier with the other. So just forget all this difficult writing here, use the next process below here. It worked very good.

 

#2. Jac's method

  • The mainframe input is at pins 7A and 7B, of the green connector inside, but I attached two wires inside the vertical module, corresponding to those pins. Of course not solder the wires on the edge connectors, but on the resistor before that. This gives me the missing measurement point. 7A and 7B are a differential input, so any hum you introduce by that will play a small role, but I do recommend to use a precise voltmeter with a floating input. Beware, THIS step is going to be the reference for the calibration precision. My Agilent 34401A is perfect for this. I expect a good quality hand held multi meter with RMS capability is able to do the same. So this is the assumption for the calibration which follows now:
  • The sensitivity of the deflection amplifiers at the backplane is 50mV/cm. So to display square wave of four divisions, we would need 200mV at pins 7A and 7B. For this, I used the external oscillator, and connect it to the vertical module input. Then adjust the red "cal" knob until exactly 200mV is present on pins 7A and 7B. Measured with the Agilent 34401Amulti meter.
  • So the calibration error of the vertical module plays no role at this moment, we only look for 200mV at pins 7A and 7B, no matter where is comes from.
  • Warm up the whole set up now for at least 1hour and re-adjust the 200mV.
  • After that, calibrate the vertical amplifier until the CRT displays four divisions. See next pictures for this.
  • Enlarge the picture, to see the settings
  • Make the dot very dim, so you don't burn the screen.
  • Set it nicely in the center
  • In the next pictures, the movements should be symmetric, otherwise you need to fix such a Geometry error first.

 

  • This is the calibration of the vertical module. I have put a 200mV peak signal on there (so 400mV peak to peak) using a square wave. With a sine wave, or triangle, you get a line, but I find it easier when it is two dots.
  • When the module is good, we get two nice dots.
  • You can move the horizontal position to the left and right with the time base position control, so to see if there is any geometry error. If so, correct this first.
  • Repeat for all four channels, and the vertical calibration is finished.
  • Not shown here, to access the modules, you need to take off the side plates of the scope, but for final check, I put them back on.
  • The 5B10 time base has an external 50mV/Div input. This bypasses the time base itself, and gives direct access to the horizontal input. Since this is 1:1 coupled to the mainframe, we have no calibration problem with this. The 1:10 attenuator in there seems precise too, so in my case I used the 0.5V input, so to have less hum. If you see however a deflection difference between the 50mV and 0.5V input, use the 50mV input.
  • I have put a 200mV peak signal on there (so 400mV peak to peak) using a square wave
  • There is nothing to calibrate here with the time base.
  • Just set the horizontal amplifier gain for 8 divisions, and that finishes this part.
  • Without signal,
  • This is a simulated "Cardiac" signal, which you can select with the Agilent 33120A Arbitrary Waveform Generator. It is just a stored "fun" example of what this instrument can do.
  • It seems just right as a time mark generator. We use it in the next picture.
  • The reference for the adjustment is the horizontal center line of the screen.
  • I did the same with a triangle wave and it works just ass good, when you position top points of the triangle exactly on the horizontal center line.
  • This is exactly the same signal as above, but enlarged, and the bottom part is not visible like that.
  • Now it looks the same as a time mark generator.
  • Here used by me at 5kHz.
  • I calibrated the speed of the time base with it.
  • Double check with the internal calibrator. It should work accurate now
  • Be careful, it has not 50% duty cycle, but frequency and amplitude should be exact
  • This signal will need another calibration of the gain of the vertical amplifier was adjusted. If so, the adjust the +30V (and -30V) slightly, bug within specification.
  • Then verify the horizontal and vertical amplifiers once more.
  • This ends the calibration.

 

DONE!